Technical Specifications
| Generator | Constant potential, microprocessor controlled, operating frequency 66 kHz | ||||||
| X-ray tube | Toshiba D-041SB | ||||||
| Focal spot size |
0.4 mm according to IEC 60336 | ||||||
| Cone diameter | 60 mm (2.36 in.) Rectangular 36 x 45 mm (1.42 x 1.77 in.) |
||||||
| Max. symmetrical radiation field | Ø60 mm at SSD 200 mm Ø60 mm at SSD 300 mm according to IEC 806 |
||||||
| Total filtration | min. 2.5 mm Al equivalent at 70 kV according to IEC 60522 |
||||||
| Inherent filtration | 1 mm Al equivalent at 70 kV according to IEC 60522 |
||||||
| Anode voltage | 60, 63, 66, 70kV | ||||||
| Anode current | 8, 7, 6, 5, 4, 3, 2 mA | ||||||
| Exposure times | 0.01–2 sec. 24 steps | ||||||
| SSD (Source-Skin Distance) Standard/Long | 200 mm (8 in.)/300 mm (12 in.) | ||||||
| Mains voltage | 100 V~/110-115 V~/220-240 V~, 50/60 Hz | ||||||
| Duty cycle | 1:13.5 | ||||||
| Electrical classification | Class I Type B | ||||||